6.4 Fraction collector F9-R test
Method description
The Fraction collector F9-R test checks the functionality of Fraction collector F9-R.
The method run takes approximately 3 minutes.
The fraction collector test is evaluated manually and no report is generated.
Note:
Required configuration
A correctly installed Outlet valve, and a correctly installed Fraction collector F9-R are
required to run the test.
Required material
The following materials are required:
Distilled water
Syringe, 25 to 30 ml
7 tubes, 2 ml or larger, for collecting the fractions.
Prepare the Fraction collector
F9-R test
Follow the instructions to prepare the system before method start.
ActionStep
Make sure that the Fraction collector F9-R is connected to port Out 10 on
Outlet valve V9-O or V9H-O
1
Direct outlet tubing W to a waste container.2
Place inlet tubing A1 into distilled water.3
Prime inlet A1 and purge System pump A. See Section 5.4.1 Prime buffer inlets
and purge system pumps, on page 166.
4
Place 7 tubes in the Fraction collector, in positions 1 to 7.5
In the System Control module, on the System menu, click Settings6
Result: The System Settings dialog box opens.
250 ÄKTA avant User Manual 29035184 AE
6 Performance tests
6.4 Fraction collector F9-R test