EasyManua.ls Logo

GE AKTA avant - Troubleshooting: Instrument Communication

GE AKTA avant
538 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
8.8 Troubleshooting: Instrument communication
Possible cause and actionScenario
Cause: Conductivity monitor cable is not connected.Error message in
UNICORN: (Error) The
first conductivity moni-
tor has been reported
lost by the instrument...
Action:
1
Switch off the instrument.
2
Connect the cable.
3
Switch on the instrument.
4
In the displayed dialog box in UNICORN, select the
option Restart the system only and click OK.
Cause: A cable to a module (including module panels) is
not connected.
Multiple error messages
in UNICORN: Lost mod-
ules.
Action:
1
Switch off the instrument.
2
Check all modules and connections.
3
Switch on the instrument.
4
In the displayed dialog box in UNICORN, select the
option Restart the system only and click OK.
The UNICORN client has lost connection to the in-
strument server during a temporary overload of the
processor
Restart the UNICORN client to regain control. The run
continues and no data will be lost.
UNICORN has lost com-
munication with the in-
strument server
Cause: The module is not functioning properly.Warning message in
UNICORN: Instrument
module is missing.
Action:
1
In the displayed dialog box in UNICORN, select the
option Restart the system only and click OK.
2
If the problem still remains, replace the module or
contact Service.
ÄKTA avant User Manual 29035184 AE 413
8 Troubleshooting
8.8 Troubleshooting: Instrument communication

Table of Contents

Other manuals for GE AKTA avant

Related product manuals