9.7.1 System settings - UV
The following table describes the UV related system settings available for ÄKTA avant.
DescriptionInstruction name
Enables or disables the alarm for the UV 1 signal from
UV monitor U9-M. When enabled, it sets the alarm limits
for the UV 1 signal. When the signal falls outside the set
limits, an alarm is issued and the method will be paused.
Note:
It is not possible to set an alarm signal for the UV 2 or
UV 3 signals from UV monitor U9-M.
Alarm UV1
Enables or disables the alarm for the UV signal from U9-
L, 2nd. When enabled, it sets the alarm limits for the UV
signal from UV monitor U9-L, 2nd. When the signal falls
outside the set limits, an alarm will be triggered and the
method will be paused.
Alarm UV 2nd
Sets the wavelengths for UV monitor U9-M. The corre-
sponding curves in the chromatogram are marked with
the set wavelengths.
Note:
The instruction is available when UV monitor U9-M
(variable) is selected in the component list. For best
performance, do not use more wavelengths than
necessary for the application.
At low wavelengths, the eluent can have significant
absorbance of its own.
Wavelength
Filters the noise in the UV signal from U9-M or U9-L
(depending on the configuration). A column-specific
averaging time is set automatically when a column is
defined in a method run and Averaging time is set as
a variable.
Noise reduction UV
Filters the noise in the UV signal from U9-L, 2nd. A col-
umn-specific averaging time is set automatically when
a column is defined in a method run and Averaging
time is set as a variable.
Noise reduction UV 2nd
460 ÄKTA avant User Manual 29035184 AE
9 Reference information
9.7 System settings
9.7.1 System settings - UV