5.5.2 Replace the Mixer module
Maintenance interval
Replace the Mixer module when the existing module malfunctions.
Required material
The following is required:
•
Mixer module (available as a spare part from GE)
•
7 mm socket tool
Procedure
Follow the steps below to replace the Mixer module.
ActionStep
Disconnect the tubing from the mixer chamber.1
Pull the mixer chamber away from the module front.2
Remove the retaining screw that holds the module in the instrument chassis
(see the Mixer Kit Installation Instruction supplied with the replacement
mixer).
3
Install the new module as described in the instructions provided with the
Mixer Kit.
4
112 ÄKTA pilot 600 User Manual 29274559 AA
5 Maintenance
5.5 Replacement procedures
5.5.2 Replace the Mixer module