Corrective actionFailed test
Make sure that the temperature compensation is turned
on (the temperature compensation factor is not zero). If
the test is performed in a cold-room, the temperature
compensation factor should be 2.1.
Curve amplitude stability
(conductivity)
The temperature compensation factor is accecssed
through System Settings:Conductivity:Cond temp
compensation.
132 ÄKTA pilot 600 User Manual 29274559 AA
7 Troubleshooting
7.3 System test failure