NOTICE
Keep UV flow cell clean. Do not allow solutions containing dissolved
salts, proteins or other solid solutes to dry out in the flow cell. Do
not allow particles to enter the flow cell, as damage to the flow cell
may occur.
NOTICE
Avoid condensation. If ÄKTA pure is kept in a cold room, cold
cabinet or similar, keep it switched on in order to avoid condensa-
tion.
NOTICE
Avoid overheating. If ÄKTA pure is kept in a cold cabinet and the
cold cabinet is switched off, make sure to switch off ÄKTA pure and
keep the cold cabinet open to avoid overheating.
NOTICE
Place the computer in room temperature. If the ÄKTA pure instru-
ment is placed in a cold room, use a cold room compatible comput-
er or place the computer outside the cold room and use the Ether-
net cable delivered with the instrument to connect to the computer.
NOTICE
UV and conductivity flow cells on the high pressure side. When
placing UV and/or conductivity flow cells on the high pressure side
of the column, the UV flow cell has a maximum pressure limit of 2
MPa (20 bar) and the conductivity flow cell has a maximum pressure
limit of 5 MPa (50 bar).
ÄKTA pure Operating Instructions 29022997 AF 25
2 Safety instructions
2.1 Safety precautions