DescriptionStateDisplay
The system has been paused
due to an alarm. To resume the
run, acknowledge the alarm and
continue the run in UNICORN.
Alarms and
errors
The Power/Communication in-
dicator displays a constant light.
and the Alarm and error indica-
tor flashes.
The system is in power-saving
mode.
Power-saveThe Power/Communication in-
dicator displays a pulsating light.
A module is being re-pro-
grammed to be compatible with
the current instrument configu-
ration.
Re-program-
ming
All indicators are lit in a wave
pattern.
ÄKTA pure Operating Instructions 29022997 AF 49