Calibration
ActionStep
For calibration, set Flow to at least 20 l/h for LowFlow Kit and 40 l/h for
HighFlow Kit.
1
Set Column under FlowPath to Inline.2
When flow signal is stable at set flow rate, select P201Gain under Pump and
click Execute.
3
Set Gradient under Pump to 100%B and wait for a stable flow signal.4
ÄKTA ready Operating Instructions 28960345 AD 105
7 ÄKTA ready gradient
7.3 Calibrate and operate the ÄKTA ready gradient pump