Valves
Problems with valves can be caused if:
SolutionProblem
To close an open valve safety lock:Valve safety lock not
closed.
1
Switch the Installation switch (in upper, right corner
of cabinet front; see Components on system cabinet
front, on page 38) to position FLOW KIT INSTALL.
2
Close the safety lock by turning the handle 180 de-
grees (see Finishing flow kit installation, on page 95).
3
Turn the installation switch back to RUN.
Note:
FLOW KIT INSTALL may fill all parts of the flow kit with
fluid from the air trap, also inlet and outlets that are cur-
rently not used. To avoid spillage of liquid, do not remove
the end caps from connections that are not used.
If part of the tubing has not been properly inserted in a
valve:
Tubing not properly in-
serted.
1
Turn the Installation switch to position FLOW KIT
INSTALL.
2
Open the valve lock.
3
Insert the tubing.
4
Close the valve lock.
5
Turn the installation switch back to RUN.
If problems remain
If problems continue after checking the points above, please contact your GE Healthcare
representative (contact information on back cover).
ÄKTA ready Operating Instructions 28960345 AD 121