ActionStep
In the Phase Properties tab, select the newly created column in the Column
type list.
1
In the Method Outline click User Defined.2
In the Phase Properties tab edit the values for the different variables (e.g.
pressure limits) to fit the column and media used. See PID control in
ÄKTA ready, on page 215 and the user documentation for the column.
Tip:
Enter values for Flow_Efficiency_Test {l/hour} and Rinsing_Flow_Rate
{l/hour} corresponding to a linear flow of 100 cm/h for the column used.
3
Tip:
Column parameter values can also be found in ReadyToProcess columns
User Manual.
ÄKTA ready Operating Instructions 28960345 AD 131