Ending
Normal completion
If no unexpected events occur during the run, UNICORN enters END state at method
completion without need for user interaction.
End before method has finished
To end the run before the method has finished, click on the End button at the top of the
Control module window.
This will produce a confirmation dialog box. Click OK in the dialog box to end the run, or
click Cancel to keep running.
In the dialog box, you can choose to save the (partial) results from the run so far. If the
run is part of a scouting run, you are also given the choice of ending this. (If you do not,
the next run will start automatically.)
Emergency stop
In case of emergency, stop the run by pressing the EMERGENCY STOP button. This will
immediately stop the pump and close all inlet and outlet valves.
To restart the system after an emergency stop, turn the EMERGENCY STOP button to
release it, and press the PAUSE/CONTINUE button.
ÄKTA ready Operating Instructions 28960345 AD 177
11 Run a method
11.2 Starting a run