ActionStep
20 25 30
77
80
85
90
95
97
When finished, wait for 15 minutes
until the temperature is constant in
the range 20°C to 30°C.
4
In the ÄKTA ready System Control
module, on the System menu, click
Calibrate. Select Cond_Cal on the
Monitor menu.
5
Read the conductivity value displayed
under Measured value and compare
it with the theoretical value from the
graph opposite at the temperature
of the calibration solution.
If the displayed value and the
theoretical value correspond, no
further action is required.
If the values differ, proceed with
step 7.
6
Enter the theoretical conductivity
value according to the graph in the
Reference value 1 field.
7
Click Read value 1. The new cell constant is automatically calculated and
updated.
8
Testing the temperature sensor
The accuracy of the temperature sensor can be checked using a Temp test tool (Sec-
tion 14.4 Ordering information, on page 210). Temperature testing is normally done once
a year, during preventive maintenance.
190 ÄKTA ready Operating Instructions 28960345 AD
12 Maintenance
12.2 Yearly maintenance