Then...When...
Select Advanced exposure.Full control of exposure
parameters is required
for special applications
Fluorescence Epi-RGB
This method is used for fluorescence imaging: single or multichannel.
Follow the instructions below to perform Fluorescence Epi-RGB capture.
ActionStep
In the Capture tab, select Fluorescence.1
Result: The options available for fluorescence capturing are displayed to the
right on the screen.
To use the full dynamic range of the image file, check High dynamic range.2
To minimize exposure time, leave High dynamic range unchecked.
Note:
This function is only effective with the Auto and Semi-Auto exposure settings.
Amersham Imager 680 Operating Instructions 29271308 AB 143
5 Operation
5.5 Capture an image
5.5.4 Fluorescence capture