Dimensions and weight
WeightDimensionsUnit
43 kg360(W) x 785(H) x 485(D) mmAmersham Imager 680
Power supply
DataParameter
100 to 240 V ACInput voltage
±10%Voltage variation
50/60 HzFrequency
250 WMax power
Environmental requirements
The following table describes the environmental requirements for Amersham Imager
680.
CAUTION
Do not use the instrument in a room with a temperature above
+28°C. Higher temperatures does not allow the CCD to cool down
properly (to -25°C).
RequirementParameter
Temperature: 18°C to 28°C (with temperature fluctuation below 10ºC
per hour or lower)
Operating temperature/humidity
conditions
Humidity: 20% to 70% RH (no dew condensation)
Note:
When the above conditions cannot be satisfied, take appropriate
actions.
Amersham Imager 680 Operating Instructions 29271308 AB 245
9 Reference information
9.1 Specifications