AquaTrans™ AT600 User’s Manual 15
Chapter 2. Clamp-On Fixture and Transducer Installation
Figure 16: Moving the Downstream Transducer
2.1.2 Transducer Spacing = 320 to 940 mm, Traverses = 2, Fixtures = 2 (cont.)
IMPORTANT: Tight contact must be made between the bars on the two fixtures to ensure an accurate spacing.
5. See Figure 17 and Figure 18 below for examples of a completed installation in the following
situations:
Note: In this drawing,
LSL is the Lower Specified Limit and USL is the Upper Specified Limit for each fixture.
a. For a calculated transducer spacing of 320 to 620 mm for a 1 MHz or 0.5 MHz transducer, locate
the upstream transducer at the
USL1 position on the first fixture. Then, locate the downstream
transducer at the calculated transducer spacing position (
USL2) on the second fixture.
Figure 17: Transducer Spacing of 320 to 620 mm with 2 Fixtures
b. For a calculated transducer spacing of 620 to 940 mm for a 1 MHz or 0.5 MHz transducer, locate
the upstream transducer at the
LSL1 position on the first fixture. Then, locate the downstream
transducer at the calculated transducer spacing position (between
LSL2 and USL2) on the second
fixture.
UPSTREAM
DOWNSTREAM
UPSTREAM DOWNSTREAM
Spacing
LSL1
USL1 LSL2
USL2
0
300
620