AquaTrans™ AT600 User’s Manual 59
Chapter 4. Initial Setup and Programming
Setting the Measurement Type
Setting the Alarm State
Select
Measurement and press []. A screen similar to the following opens.
In the
Measurement menu, select the type of analog output to be used, and
press
[]. The screen returns to the previous display.
Use the [] or [] arrow keys to select
Alarm State and press[]. A screen
similar to the following opens.
Two alarm states are available:
• Normal: Normally open, the alarm contacts close in an error condition
• Fail Safe: Normally closed, the alarm contacts open upon in an error
condition or a power failure
To change the state of the alarm, select the desired status and press
[],.
The screen returns to the previous display.
Alarm
Measurement
Alarm State
Alarm Type
Alarm Value
Velocity
Act Volumetric
Std Volumetric
Mass
Alarm
Measurement
Alarm State
Alarm Type
Alarm State
Normal
Fail Safe