September 2004
2-10 Programming Site Data
Fluid Type 1. The selections for fluid type vary depending on whether the
Tracking Window is enabled or disabled. Press [
W] and [X] to
select the desired fluid and press [ENTER]. Refer to Table 2-2
below for a list of available fluids.
Table 2-2: Fluid Types
2. If you selected OTHER, enter the additional information as
prompted by using the arrow keys. Press [
W] and [X] to move the
cursor to desired location and press
[S] and [T] to increase or
decrease the number and press [ENTER].
Reynolds Correction 1. Press [W] and [X] to select whether to activate or deactivate the
Reynolds Correction Factor and press [ENTER].
Note: Typically, Reynolds Correction is set to active.
2. Do one of the following:
• If you selected OFF, enter the Calibration Factor and press
[ENTER]. Then, proceed to one of the following sections:
• For Clamp-on Transducers - proceed to Number of
Traverses and Transducer Spacing on page 2-12.
• For Wetted Transducers - The meter returns to the Channel
PROGRAM window. At this point, you can continue
programming in this submenu as desired. To leave the
Channel PROGRAM window, press [ESC] three times. To
complete setting up the meter, you must enter data in the
GLOBL-SYSTM submenu as described on page 2-22.
• If you selected ACTIV, proceed to KV Input Selection on
page 2-11.
Tracking Windows =
NO YES
WATER W100 (Water, 0°-100°C)
OTHER W260 (Water, 0°-260°C)
Oil, Tracking
OTHER