September 2004
4-2 Calibration
Accessing the Calibration
Menu
1. Press [ESC], [ENTER], [ESC]. The meter enters the user program.
2. Press [
X] until CALIB appears and press [ENTER].
3. Press [
X] until SLOT0 appears and press [ENTER].
4. Press [
X] until A or C appears and press [ENTER].
Calibrating the Low End of
the Analog Output
1. Press [X] to select 4 mA and press [ENTER].
2. Press [
X] to select UP or DOWN to adjust the 4 mA reading and
press [ENTER].
3. When the desired reading is achieved, select STORE. If the
ammeter reading cannot be adjusted with 5.0 µA of the 4 mA
setting, select ABORT to end calibration and contact the factory for
assistance.
Calibrating the High End of
the Analog Output
1. Press [X] to select 20 mA and press [ENTER].
2. Press [
X] to view the options to adjust the 20 mA reading and
press [ENTER].
3. When the desired reading is achieved, select STORE. If the
ammeter reading cannot be adjusted with 5.0 µA of the 4 mA
setting, select ABORT to end calibration and contact the factory for
assistance.
Procedure Options After completing the above steps, the meter returns to the ANALOG
OUTPUT window. Do one of the following:
• Proceed to Step 1 under Testing Linearity on the next page.
• To perform another programming function, refer to Appendix
A, Menu Maps, to navigate to the desired menu.
• To leave the User Program, press [ESC] four times.