Appendix B. Using the Clamping Fixtures
106 PanaFlow™ LC User’s Manual
Figure 94: UCF Setup, Odd Traverse, Step 9
11. Use the screw hooks to tighten the chains on both the fixed and adjustable short blocks until both
blocks are tightly secured to the pipe.
IMPORTANT: Make sure the chains are perpendicular to the clamping fixture and are not twisted. If the chains
are slanted, the slack may cause the fixture to move. Also, the transducer spacing dimension may
change after the transducers are mounted.
Figure 95 below shows a completed odd traverse installation without the transducers. Proceed to the section
on mounting the transducers later in this chapter.
Figure 95: A Odd Traverse Clamping Fixture Installation without Transducers