PanaFlow™ LC User’s Manual 83
Chapter 4. Error Codes and Troubleshooting
S6: Flow Measurement One or more flow
measurement channels are in
error
Error: One or more flow measurement channels
are in error; accuracy of the measurement may
be compromised. For more details please check
flow(E) errors
S7: Persistent Param
CRC
Persistent parameter CRC
fault
Fault: Persistent parameter CRC failed. Try
power cycling the meter. If error persists after
power cycle, contact BHGE factory
S11: Clock Frequency Clock frequency error Fault: Input clock frequency failure. Try power
cycling the meter. If error persists after power
cycle, contact BHGE factory
S12: CPU CPU error Fault: CPU registers have stuck bits. Try power
cycling the meter. If error persists after power
cycle, contact BHGE factory
S13: Invariable Flash
Memory
Flash memory fault Fault: Flash memory test failed. Try power
cycling the meter. If error persists after power
cycle, contact BHGE factory
S14: Invariable SRAM Invariable SRAM fault Fault: Invariable SRAM memory test failed. Try
power cycling the meter. If error persists after
power cycle, contact BHGE factory
S15: Variable Memory Variable SRAM fault Fault: Variable SRAM test failed. Try power
cycling the meter. If error persists after power
cycle, contact BHGE factory
S16: FPGA Config FPGA configuration error Fault: FPGA configuration validation failure. Try
power cycling the meter. If error persists after
power cycle, contact BHGE factory
S17: Temperature Temperature error Error: Temperature of the electronics is outside
the pre-defined operating range. Make sure that
the ambient temperature is not outside the
meter operating range
S18: Driver Fault Driver failure Fault: Driver failure. Try power cycling the
meter. If error persists after power cycle, contact
BHGE factory
S19: Watch Dog Failure Watch dog failure Fault: Watch dog test failed. Try power cycling
the meter. If error persists after power cycle,
contact BHGE factory
S21: Stack Overflow Stack overflow Fault: Stack overflow. Try power cycling the
meter. If error persists after power cycle, contact
BHGE factory
S22: Sequence or
Window Watchdog
Sequence failed Fault: Sequence failure detected. Try power
cycling the meter. If error persists after power
cycle, contact BHGE factory
Table 24: System Error Description and Recommended Actions
Error Code Error Message Description / Recommended Action