Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
3-30
box is checked.
7. Query & Retrieve Setup and
Check
Query & Retrieve setup:
Click Query & RetrieveTab on
setup screen
In local settings, setup local
transport parameters:
Local Listening Port
Time Out
Days to Cache Retrieved
Images
Click “Add…” button into Add
interface. Fill in the information
of the server of Query &
Retrieve.
Click “Edit…” button into Edit
interface, edit the information of
the server of Query & Retrieve.
In Add and Edit interface, click
“Verify” button. The message
“The DICOM server has verified
successfully.” will display if the
server has connected. Click “OK”
button and then select the “Exit”
button to close Add or Edit
interface.
Add and Edit Interface of Query & Retrieve