Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-22
ID
Warning
Description
(Displayed in User Interface)
memory failed
Image acquisition exception.
System can’t take exposure.
The mainframe stopped.
Please restart the system. If
this issue persists, please call
for service.
1. If Gige camera is not
working correctly, please
change camera.
2. If LED on ethnet interface
is not twinkle, please insert
the network cable into
ethnet interface again or
replace another network
cable.
3. If all above methods can’t
fix the exception, please re-
template failed
Shading parameter loading
failure. Image quality may be
impacted. Please restart the
system. If this issue persists,
deployed corrected.
template failed
exception. The parameter was
not saved. Please re-calibrate
Disk is full or directory has
no permission to write.
calibration
failure
Load shading calibration data
failure. Image quality may be
affected. Please restart the
system. If this issue persists,
memory failed
Run out of memory. Please
restart the system. If this issue
persists, please call for service.
failed
Failed to find the display
device. Images can’t be
displayed. Please restart the
system. If this issue persists,
connection, if display driver
installed correctly
context failed
Display driver exception. Image
can’t be displayed. Please
restart the system. If this issue
persists, please call for service.
installed correctly
Create display
window failed
Display driver exception. Image
can’t be displayed. Please
make sure display driver
installed correctly