Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-24
ID
Warning
Description
(Displayed in User Interface)
reason list
failed
failure. Abandon Exam
function will be impacted.
Please restart the system. If
this issue persists, please call
for service.
workstation
2. Check display card of
workstation
3. Reinstall workstation
software.
printer list
failed
Connectivity configuration
loading failure. DICOM printer
list missed. Please restart the
system. If this issue persists,
please call for service.
2. If not work, re-install
Workstation software.
configuration
file failed
Setup configuration loading
failure. Some setting may be
missed. Please restart the
system. If this issue persists,
please call for service.
2. If not work, re-install
Workstation software.
option failed
Setup configuration loading
failure. Exam list display setting
missed. Please restart the
system. If this issue persists,
please call for service.
MPPS server
failed
Connectivity configuration
loading failure. Default MPPS
server setting missed. Please
restart the system. If this issue
persists, please call for service.
2. If not work, re-install
Workstation software.
storage server
failed
Connectivity configuration
loading failure. Storage server
setting missed. Please restart
the system. If this issue
persists, please call for service.
2. If not work, re-install
Workstation software.
title failed
Connectivity configuration
loading failure. Local AE title
missed. Please restart the
system. If this issue persists,
please call for service.
2. If not work, re-install
Workstation software.
server list failed
Connectivity configuration
loading failure. Worklist server
list missed. Please restart the
system. If this issue persists,
please call for service.
2. If not work, re-install
Workstation software.