Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-24
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0306505
Warning
Load abandon
reason list
failed
Abandon Reason loading
failure. Abandon Exam
function will be impacted.
Please restart the system. If
this issue persists, please call
for service.
1. Check mainboard of
workstation
2. Check display card of
workstation
3. Reinstall workstation
software.
0306506
Warning
Load DICOM
printer list
failed
Connectivity configuration
loading failure. DICOM printer
list missed. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0306507
Warning
Load setup
configuration
file failed
Setup configuration loading
failure. Some setting may be
missed. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0306508
Warning
Load exam list
option failed
Setup configuration loading
failure. Exam list display setting
missed. Please restart the
system. If this issue persists,
please call for service.
0306509
Warning
Load default
MPPS server
failed
Connectivity configuration
loading failure. Default MPPS
server setting missed. Please
restart the system. If this issue
persists, please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0306510
Warning
Load DICOM
storage server
failed
Connectivity configuration
loading failure. Storage server
setting missed. Please restart
the system. If this issue
persists, please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0306511
Warning
Read local AE
title failed
Connectivity configuration
loading failure. Local AE title
missed. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.
0306513
Warning
Read MWL
server list failed
Connectivity configuration
loading failure. Worklist server
list missed. Please restart the
system. If this issue persists,
please call for service.
1. Restart the system.
2. If not work, re-install
Workstation software.