Loading...
Troubleshooting
6-49
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0411518
Warning
Create image
file failed
Image file create failure. Please
restart the system. If this issue
persists, please call for service.
No permission or no
enough space on disk
0411519
Warning
Move image file
failed
Image file move failure. Please
restart the system. If this issue
persists, please call for service.
File crashed or no
permission
0411520
Error
IM module
register servant
exception
Workstation communicaiton
exception. Please restart the
system. If this issue persists,
please call for service.
make sure if omniNames
service failed, any module
initialization failed
0411522
Error
IM module
unregister
servant
exception
Workstation communicaiton
exception. Please restart the
system. If this issue persists,
please call for service.
make sure if omniNames
service failed, check
omniOrb log file
0411527
Error
IM module
resolve object
exception
Workstation communicaiton
exception. Please restart the
system. If this issue persists,
please call for service.
make sure if omniNames
service failed, any module
crashed
0411528
Error
IM module
Invoke corba
service
exception
Workstation communicaiton
exception. Please restart the
system. If this issue persists,
please call for service.
make sure if omniNames
service failed, any module
crashed
0411529
Warning
FIFO is full
Image FIFO full. Image display
refreshing may be slow. Please
restart the system. If this issue
persists, please call for service.
Check CPU Affinity and
thread priority. Complied
SW version (Debug or
Release)
0411531
Warning
Create
directory failed
System file operation failed.
Please restart the system. If
this issue persists, please call
for service.
file crashed or no
permission
0411532
Error
IM module
syncNotify
event exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0411533
Error
IM module
asyncNotify
event exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.

Table of Contents