6-50
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
0411801
Warning
System's image
capacity reach
limit
The image can’t be saved.
Image storage capacity limit of
100,000 images reached.
Please delete useless or out-of-
date images to make disk
space for new images.
remove some images
0411802
Warning
Exam's image
capacity reach
limit
Image storage capacity limit of
9999 images for each exam
reached. No more images can
be saved under the exam.
Please delete some useless
images if you want to save
new images of the exam.
0403501
Error
DTR sets locale
failed
System start failure. Set locale
failed. Please restart the
system. If this issue persists,
please call for service.
0403502
Error
DTR reports
ready failed
System start failed. Please
restart the system. If this issue
persists, please call for service.
0403503
Warning
DTR parse
retrieved image
failed
Parse retrieved image failure.
Please try again. If this issue
persists, please call for service.
0403504
Warning
DTR gets exam
information
failed
Parse exam information of raw
images failure. Please try
again. If this issue persists,
please call for service.
0403505
Warning
DTR creates
RDSR object
failed
Create DICOM RDSR failure.
Please try again. If this issue
persists, please call for service.
0403506
Warning
DTR creates
DICOM file
setup failed
Create DICOM file exception.
Please try again. If this issue
persists, please call for service.