Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-52
Exception
ID
Error
Warning
Exception
Description
(In system Log)
Message
(Displayed in User Interface)
Diagnosis Reference
exception
0403518
Warning
DTR imports
raw image
teardown
exception
Import raw image exception.
Please try again. If this issue
persists, please call for service.
0403519
Error
DTR module
resolve object
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0403520
Error
DTR module
Invoke corba
service
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0403521
Error
DTR module
register servant
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0403522
Error
DTR module
unregister
servant
exception
Workstation communication
exception. Please restart the
system. If this issue persists,
please call for service.
0405501
Warning
SCU prints
image failed
DICOM print failure. Please
make sure the network and
DICOM printer configuration
are correct, and the printer is
available. And try again. If this
issue persists, please call for
service.
Check DICM configuration.
0405502
Warning
SCU performs
echo operation
failed
Verify the DICOM server failure.
Please make sure the network
and DICOM server
configuration are correct, and
the server is available. And try
again. If this issue persists,
please call for service.
Check DICM configuration.