Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
4-26
Camera Centering Calibration procedure for Brivo OEC 865:
1. To have access to the Camera Securing Screws, perform the following.
a. Position the C-arm to Image Intensifier up, Monoblock down.
b. Remove the I.I. back cover and cables connected to Camera.
c. Remove the upper counter balance
d. Release 4 secure screws by TORX T20 screw driver.
2. Log in software service mode, in calibration TAB, click “Beam alignment…” button
3. Use 2 camera alignment screws (by 1/8 Inch HEX wrench) to adjust the camera center. (On the
other side are two spring screws which do not need to be adjusted, just fasten/loose alignment
screws can center the camera.) Perform camera centering check and repeat adjusting camera
alignment screws as many times as necessary.