Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
6-8
ID
Warning
Description
(Displayed in User Interface)
Generator power exception.
Exposure is disabled. Please
restart the system. If this issue
persists, please call for service.
1. Check input voltage and
power supply.
2. Check DC Bus fuse and
related cables.
3. Check the IGBT.
4. Check the KV control
board.
Generator power exception.
Exposure is disabled. Please
restart the system. If this issue
persists, please call for service.
between the kV control
board and IGBT board.
2. Check the IGBT.
3. Check the KV control
board and power supply to
120KV
Generator exception. Exposure
is disabled. Please restart the
system. If this issue persists,
35kv
Generator exception. Exposure
is disabled. Please restart the
system. If this issue persists,
please call for service.
1. Check input voltage and
power supply.
2. Check DC Bus fuse and
related cables.
3. Check the IGBT.
4. Check the KV control
board.
5. Redo KV calibration.
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1. Check the bundle cable.
2. Check KV_feedback
signal on kV control board.
current by HW
monitor
Execution exception. Please
restart the system. If this issue
persists, please call for service.
1. Check DC power supplier
and fuse F6 on the Mother
board (backplane).
2. Check related cables.
3. Check the MCB.
4. Check the FDB.
current
Execution exception. Please
restart the system. If this issue
persists, please call for service.
2. Check DC power supplier
and fuse F5 on the Mother
board (backplane).
2. Check the FDB.
3. Replace the monoblock.