Brivo OEC 715/785/865 Mobile C-Arm X-Ray Product Service Manual
7-64
Interconnect Cable Socket kit: (PN#: 5075972)
Interconnect Cable Socket Kit (FRU PN: 5075972) includes Interconnect cable and Interconnect
Cable socket. For existed IB system, FSE should replace the interconnect cable socket and
upgrade the interconnect cable(if Rev.13 or lower version).
Connector List:
Emergency Switch MF_AC_28V_N
Emergency Switch
Emergency Switch MF_AC_28V_L
To Mainframe Mother Board
Replacement procedure:
1. Remove Mainframe back and front covers refer to section 7.
2. Cut all the cable ties and unplug the connectors
5085708 Interconnect cable
5085709 Interconnect Cable socket
5075972
Interconnect
Cable Socket