Brivo OEC 715/785/865/865 Mobile C-Arm X-Ray Product Service Manual
2-30
of interlock loop
MCB as part of the whole interlock
loop. If connecter J2 of MCB is
loose, the whole inter loop will be
broken, Mainframe will power off.
Test Points
KV control signal used to calibrate mA feedback
KV control signal to KV control board
mA feedback signal without KV calibration
mA sample signal with KV calibration
mA control signal to Filament board
Output of proportional & integral circuit
mA control signal without PREHEAT
mA control signal offset setting
Filament control signal from FDB
LED Indicators
Light lights when +2.5V is available
Light lights when +1.2V is available
RXD485+ & RXD485-. D20 on MCB lights normally, and flickers in
communication
KEYSW_OFF & KEYSW_ON & KEYSW_COM. D22 lights when "KEYSW_ON" & turns
off when "KEYSW_OFF"
SPI_CC_CS. D24 on MCB lights normally, and turns off when CCB is drived
FOOT_HLF & FOOT_COM. D26 turns off when FOOT_HLF & FOOT_COM are
connected
KEYSW_OFF & KEYSW_ON & KEYSW_COM. D27 lights when "KEYSW_ON" & turns
off when "KEYSW_OFF"
D41 turns off when footswitch interupt signal is available to MCU
SPI_SIMO. D29 flickers on MCB
THERMAL_SWTICH_A & THERMAL_SWITCH_B.D30 on MCB lights normally, and
turns off in overheat error
D32 turns off when FPGA configuration completes
HAND_HLF & HAND_COM. D33 turns off when HAND_HLF & HAND_COM are
connected
FOOT_BUTTON1 & FOOT_COM. D34 turns off when press
WK_A & WK_B. D35 on MCB turns off normally, and flickers in communication