Start Up and Operating Control
Page 4-10
NOTE
X-Ray Off Indicator
When the key switch is placed in X-Ray off position ( ) or standby position ( ), X-Ray is disabled
and this indictor will light.
In the other X-Ray off case, this indictor will light too.
Generator Control
kVp
Press the kVp key can adjust kVp manually and override auto technique setting. Press the top portion of
the key to increase kVp or press the bottom portion of the key to decrease kVp.
mA/mAs
Press the mA/mAs key can adjust mA/mAs manually for fluoroscopy/film respectively and override auto
technique setting. Press the top portion of the key to increase mA/mAs or press the bottom portion to
decrease mA/mAs.
Manual mode can be selected directly by pressing kVp” or mA/mAs key during auto mode.
Auto Technique
Pressing the Auto Technique key enables the system to produce an optimum image by adjusting the
technique parameter (kVp, and mA) automatically. When selected, the LED will illuminate indicating Auto.
Deselecting auto will extinguish the LED and the kVp and mA value can be adjusted manually.