Maintenance
3.Todisinfect,wipewithasoft,lint-freeclothmoistenedwithanappropriate
disinfectant.
Usethefollowingsolutions,asrecommendedintheAPICGuidelinesforSelection
andUseofDisinfectants(1996):
•Sodiumhypochlorite(5.2%householdbleach)minimum1:500dilution
(minimum100ppmfreechlorine)andmaximum1:10dilution.
•Anysodiumhypochloritewipeproductthatmeetsthepreviousguidelines
canbeused.
4.Allowthecleaningsolution/disinfectanttoremainonthedeviceforaminimum
ofoneminute,orperhospitalguidelines.
5.Wipeoffthecleaningsolution/disinfectantwithaclean,moistenedcloth.
6.Drythoroughlywithadry,lint-freeclothandletairdryforaminimumof30
minutesbeforeuse.
DONOTuseexcessivedryingtechniques,suchasoven,forcedheat,orsun
drying.
NOTE:
Dryingtimesmayvarybasedontheenvironmentalconditions.
Cautions
•Followthecleaninginstructionsexactly.
•Wringexcessdisinfectantfromwipebeforeusing.
•Neverimmersethedevice,cables,orleadwiresinanyliquid,asthismaycorrode
metalcontactsandaffectsignalquality.
•Donotallowuidtopoolaroundconnectionpins.Ifthishappens,blotdrywitha
soft,lint-freecloth.
•Neveruseconductivesolutionsorsolutionsthatcontainchlorides,wax,orwax
compoundstocleanthedevice,cables,orleadwires.
•NeverusesolutionsorproductsthatcontainanytypeofAmmoniumChloridesuch
as,butnotlimitedto:
•DimethylBenzylAmmoniumChloride
•QuaternaryAmmoniumChloridesolutions
•Abrasivecleanersorsolventsofanykind
•Acetone
•Ketone
•Betadine
•Alcohol-basedcleaningagents
•Sodiumsalts
•Neverautoclaveorsteamcleanthedevice,cables,orleadwires.
•Donotuseuntilthoroughlydry.
2049751-001C
CAM-14
®
/CAM-HD
®
15