Maintenance
Tools
1.ECGSimulator
2.Standardhandtoolsincludinga#6Torxdriver
3.Currentleakagetester
4.Hipottester
5.Anti-staticwriststrap
6.Applicableserviceand/oroperatormanualasneededforreference
VisualInspection
Inspectthefollowingforexcesswearand/oranyvisualsignsofdamage.
1.Inspectfordefectiveorbrokenpatientcable/leadwiresorout-of-dateelectrodes.
2.Reviewelectrodeplacement,skinprep,andpatientrelatedrequirementswith
theECGTech.
3.Inspectexternalsurfaces.
OperationalChecks
1.Completepower-upself-test.
2.Runasimulatedrecordedrhythmstrip.
3.RunasimulatedrecordedECG.