2002938-002E Cardiolab® II Plus Amplifier Manual 2-5
Front View
Table 4. Front View 64/32 Channels- CardioLab II Plus Amplifier
Name Description
Pressure Inputs Four pressure inputs will each accept a pressure transducer
that is calibrated to 5uV/V/mmHg. See the Accessory Parts
List for GE Medical Systems - Information Technologies
recommended pressure transducers.
ECG Input The ECG input connector will accept a 10 lead ECG input
cable.
Stimulator Input The stimulator input will accept a stimulator input cable. Use
the 3 ft., 15 ft. or 30 ft. Touchproof Stimulator Cables (part
numbers , 301-00204-03, 301-00204-15, 301-00204-30,
respectively) with the CardioLab II Plus Amplifier.
Catheter Inputs Up to three catheter inputs will each accept a 32 pole catheter
input module. Use the CardioLab II Plus Catheter Input
Modules (part number 301-00202-08) with the CardioLab II
Plus Amplifier.