EasyManua.ls Logo

GE Cardiolab II Plus - Page 33

GE Cardiolab II Plus
70 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Equipment Overview
2002938-002E Cardiolab® II Plus Amplifier Manual 2-7
Table 5. Front View 128/96 Channels- CardioLab II Plus Amplifier
Name Description
Pressure Inputs Four pressure inputs will each accept a pressure transducer
that is calibrated to 5uV/V/mmHg. See the Accessory Parts
List for GE Medical Systems - Information Technologies
recommended pressure transducers.
ECG Input The ECG input connector will accept a 10 lead ECG input
cable.
Stimulator Input The stimulator input will accept a stimulator input cable. Use
the 3 ft., 15 ft. or 30 ft. Touchproof Stimulator Cables (part
numbers , 301-00204-03, 301-00204-15, 301-00204-30,
respectively) with the CardioLab II Plus Amplifier.
Catheter Inputs Up to seven catheter inputs will each accept a 32 pole catheter
input module. Use the CardioLab II Plus Catheter Input
Modules (part number 301-00202-08) with the CardioLab II
Plus Amplifier.
Auxiliary Inputs The auxiliary input will accept the Auxiliary Input Cable (part
number 301-00205-08).

Table of Contents

Related product manuals