Equipment Overview
2002938-002E Cardiolab® II Plus Amplifier Manual 2-9
Internal Assembly
Board Identifications
Control Board (PN: 2002344-001)
Provides communications between the amplifier and the CPU.
Transmits all real-time data.
Receives all amplifier setup information from the CPU and sends
commands to other boards
Table 6. Back View - CardioLab II Plus Amplifier
Name Description
FDDI Connection to
Computer
This connector will accept a FDDI type fiber optic cable. The
fiber optic cable allows for two-way communication between
the amplifier and the CardioLab system.
Power Input The GE-supplied power cord connects to the Power Input.
Equipotentiality
Terminal
The Equipotentiality Terminal is connected to earth ground.