Installation checkout
117
2106778-001
8.3.7 Alphanumeric keyboard
Perform this test only if an alphanumeric keyboard is connected to the patient monitor.
1. Select Data & Pages > Admit/Discharge (or Start / End Case).
2. Select the Patient tab > Edit Name & MRN.
3. Press
Enter to highlight the Medical Record Number field.
4. Type some text into the Medical Record Number field using the connected alphanumeric
keyboard. Include some characters that are specific to the chosen keyboard locale to
verify that the keyboard language configuration is correct.
5. Select to close the open menu and return to the main display.
8.3.8 Barcode reader
Perform this test only if a barcode reader is connected to the patient monitor.
1. Select Data & Pages > Admit/Discharge (or Start / End Case).
2. Select the Patient tab.
3. Scan a test barcode that is applicable to your system:
8.3.9 MC Network and S/5 Network
Perform the following test only if the patient monitor is connected to a wired MC Network or S/5
Network.
NOTE: Make sure that at least one other patient monitor is on the network. The other patient
monitor must be in an admitted state and have an active ECG measurement with a simulator
signal.
Parser Type Test Procedure
Length Delimited or Character
Delimited Parser
1.
2.
3.
4. Select Scan from Barcode.
5. Scan a known test barcode obtained from the hospital.
NOTE: The barcode data content must be known and in
compliance with the completed parser configuration.
6. Verify that the data content in the barcode is correctly
populated to the related fields in the Patient and the
Administr. Information tabs.
No parser
1.
2.
3.
4. Select Edit Name & MRN and press
Enter to highlight
the Medical Record Number field.
5. Scan a sample barcode that only contains one piece of
information (e.g., a Serial Number barcode from a
module’s device label).
6. Verify that the data is correctly populated into the
Medical Record Number field.
7. Select to close the open menu and return to the main display.