Impedancerespiration
ProblemSolution
Respirationdetectionisnotdependentonthesizeof
thewaveform.Sizeisforvisualpurposesonly.
Whyisthemonitordetectingcardiacartifactas
breaths?
Thebreathdetectionthresholdistoolow(1=breath,
2=artifact).Thefollowingisanexampleofincorrect
detection,wherethe1’sarecorrectlydetected
breaths,butthe2’sareartifactincorrectlydetected
asbreaths,ascanbeseenbytheinspirationand
expirationmarkersonthe2’s.
●Increasethedetectionsensitivitypercentageuntil
themarkerscorrectlyidentifyeachinspirationand
expiration.Thefollowingisanexampleofcorrect
detection,wheretheinspirationandexpiration
markersonlyappearonthebreathsandnotthe
artifact.
132CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017