SpO₂
WARNING
Manyfactorsmaycauseinaccuratereadingsandalarms,
decreasedperfusion,andorlowsignalstrength:
Interferingsubstances:
CarboxyhemoglobinmayerroneouslyincreaseSpO
2
reading.
Methemoglobin(MetHb)usuallyrepresentslessthan1%
ofthetotalHb,butinthecaseofmethemoglobinemia
thatcanbecongenitalorinducedbysomeIVdyes,
antibiotics(suchassulphas),inhaledgases,etc.,this
levelincreasessharplyandthuscancauseinaccuracies
intheSpO
2
reading.
Intravasculardyes(suchasindocyaninegreen,
methyleneblue,etc.)
Physiologicalcharacteristics:
Cardiacarrest
Hypotension
Shock
Severevasoconstriction
Severeanemia
Hypothermia
Venouspulsations
Darklypigmentedskin
Ventricularseptaldefects(VSDs)
Nailpolishorarticialnailsatthemeasurementsite
Environmentalconditions:
Electromagneticinterference
Excessiveambientlight
Electricalinterference
Electrosurgery
Debrillation-Maycauseinaccuratereadingforashort
amountoftime.
Excessivepatient/sensormotion.Artifactcansimulate
anSpO
2
reading,sothatthedevicefailstosoundan
alarm.Inordertoensurereliablepatientmonitoring,the
properapplicationoftheprobeandthesignalquality
mustbecheckedatregularintervals.
Sensorplacement:
Incorrectsensorplacement-prolongedmonitoringor
incorrectsensorapplicationcancauseskinirritationor
impairedcirculation.Itisrecommendedthatyoucheck
theprobesiteeveryfourhours(morefrequentlyforpoor
perfusionorforneonates).Refertotheinstructions
suppliedwiththesensor.
Sensorplacementonthesameextremityasablood
pressurecuff,arterialcatheterorintravascularline;or
arterialocclusionproximaltothesensor.
2094480-001CARESCAPEONE135
CARESCAPE ONE User Manual DRAFT 26 October 2017