Invasivepressures
zeroedarterialpressure:ArtX,FemX,orUACX.Ifnoarterialpressureisavailable,the
numericallyrstzeroedpressureisused.
Triggeringintra-aorticballoonpumps
NOTE
Ifyouchoosetousethemonitorfortriggering,usethe
followinginstructions.Failuretofollowtheseinstructionsmay
resultinanincompatibleanalogoutputsignal,whichmay
contributetopatientinjury.
1.Contacttheballoonpumpmanufacturerforinterfacerequirements.Seethe
technicalspecicationsfortheECGanalogoutputdelayspecicationforthe
acquisitiondevice.
2.ConnectacompatibleanalogoutputcabletotheDebSyncconnector.
3.Adjusttheinvasivepressurelter.Ifarterialpressureisusedtotriggertheballoon
pump,usethe40Hzpressurelter.
4.PrimarydisplayedECGlead:IftheballoonpumptriggersofftheRwaveofthe
ECG,reviewthepatient’sECGleadsandplacetheonewiththegreatestamplitude
inthetop(primary)positiononthemonitordisplay.
5.Pacemakerdetection:Ifthepatienthasapacemaker,besurethepacemaker
detectionisturnedon.Failuretoturnpacemakerdetectiononmaycausepoor
beatdetection,whichmayresultininadequatetriggeringoftheballoonpump.
EffectsofIABPondisplayedvalues
Displayedpressurevaluesareaffectedbytheintra-aorticballoonpump.TheIABP
programdisplaysthreevalues,forexample150/45(98).Therstvalueisthesystolic
value,thesecondisthediastolicvalue,andthethirdisthemeanvalue.Thedisplayed
numericvaluesareindicatingarapidlyvaryingwaveformgeneratedduringIABP
treatmentanddonotalwaysreectatruearterialbloodpressure.
Foraccuracyandreliability,combinetheserecommendedmethodsforreading
arterialand/orfemoralbloodpressure:
●TheIABPwaveformdisplayedonthescreen(usescalesforevaluation)
●Thedisplayontheballoonpumpdevice,ifavailable
SincethereareanumberofpointsalongtheIABPwaveformthatcouldbethe
displayedvalue,itisimportanttoknowwhichpointstheprogramuses.Thevalues
displayedwilldifferdependingonthetimingofthepump.
1:1or1:2timing
Diastolicnumerics:Thedisplayeddiastolealwaysequalstheballoonenddiastole.
Systolicnumerics:
●Whentheaugmenteddiastoleisgreaterthanthepatientsystole,thedisplayed
systoleequalstheaugmenteddiastole.
174CARESCAPEONE2094480-001