CO₂
ProblemSolution
Whycanweseedipsinthecapnogramduring
expiration?
Dipsinthecapnogramcanbeduetopatient
conditions.“Clefts”inthecapnogramareseen
asactionofmusclerelaxantisimpactedby
spontaneousventilation.
Thedipsseeninthecapnogramduringexpiration
mayberelatedtothesidestreamgassampling,the
continuousgasowtotheY-piece,andpatient’s
cardiaccontractions,whichcauseintra-thoracic
pressurechangesandthereforeowvariations.
ApressuretransducerintheLoFloSidestream
Moduleminimizesowvariationsbymeasuring
andadjustingforchangesinairwaypressures.
WhyistheEtCO
2
valueconsiderablylowerthan
theCO
2
partialpressuredeterminedbybloodgas
analysis?
EtCO
2
valuesarealwayslowerthanpaCO
2
values
measuredbybloodgasanalysis.
Thetypicalclinicalreasonsaredead-space
ventilation,ventilation/perfusionmismatch,adrop
incardiacoutput,alveolarshunts,andincomplete
emptyingofthealveoli.
TheEtCO
2
-paCO
2
gradientcanbeanindicatorof
aclinicalconditionsuchasanincreaseinalveolar
deadspace.
Alsochecktheintegrityofthebreathing
circuit;blood-gasanalysiscorrectedtoalower
temperatureincaseofhypothermia.
Thesamplecellisinserted,butthepumpdoesnot
turnonandtheCheckAdaptermessageisdisplayed.
Checkthatthesamplecellisfullyinserted.
TheLoFloSidestreamModulemayneedtowarm
uptobedetected.
CheckthattheLoFloSidestreamModuleis
connectedtotheCARESCAPECO
2
.
Checkthesamplecelloftheaccessoryfordamage
orocclusion.