Messages
Message
Location
PossiblecausesSuggestedactions
ECGmeasurements
removed
al.areaCARESCAPEECGhasbeen
removed.
Checkallconnectionsand
reconnectasrequired.
FrequentPVCs
al.areaPhysiologicalalarm.Checkthepatientstatus.
FrequentSVCs
al.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
Tachy
HRhigh
HR(ECG)high
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Physiologicalalarm.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
IdenticalECGdeviceal.areaTherearetwoCARESCAPEECGs.
RemoveoneCARESCAPEECG.
Irregularal.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
Leadoff
LA/Lleadoff
LL/Fleadoff
RA/Rleadoff
RL/Nleadoff
V2/C2leadoff
V3/C3leadoff
V4/C4leadoff
V5/C5leadoff
V6/C6leadoff
V/Cleadoff
Va/Caleadoff
Vb/Cbleadoff
al.area,
wavef.
Oneormoreleadwiresor
electrodesdisconnected.Other
ECGleadsareavailablefor
arrhythmiadetection.
Checktheleadwireand
electrodeconnections.
Leadchangedwavef.Themonitorautomatically
switchestheECG1waveform
selectiontoameasurableECG
Lead(I,II,III,aVR,aVL,aVF,
V1,V2,V3,V4,V5orV6)ifthe
currentECG1waveformisnot
measurable.Thismessagealso
appearswhentheuserchanges
theECG1leadselection.
NotethattheECGwaveform
willbefromtheleadselected
fortheECG1lead.Checkthe
lead.
Leadsoffal.area,
wavef.
Oneormoreoftheconnected
leadwiresisdisconnectedand
arrhythmiadetectionisnot
possible.
Checktheconnections.
Learning
wavef.ECGalgorithmisinlearning
phase,messageshowne.g.
whenECGmeasurementis
started.
Noactionrequired.
MissingBeat
al.area,
wavef.
Physiologicalalarm.Checkthepatientstatus.
2094480-001CARESCAPEONE211
CARESCAPE ONE User Manual DRAFT 26 October 2017