Messages
Message
Location
PossiblecausesSuggestedactions
●Artifact
●param.
Artifactdetected.●Checksensorcontact.
●Repositionorreplacesensor.
●Requestthepatienttoremain
still.
●Checkdevice
●SpO2malfunction
●param.
●al.area
CARESCAPESpO
2
device
malfunction.NoSpO
2
data
detected.
●ChecktheCARESCAPESpO
2
device.
●Replaceifneeded.
●Checkprobe
●CheckSpO2probe
●param.
●al.area
ThereisnodetectableSpO
2
signal,thesensorisfaultyoris
detachedfromthepatient.
WithCARESCAPESpO
2
–Nellcor
thismessagealsoappears
whenthesaturationupdates
exceed30seconds,thepulse
rateislost,orthepulsetimeout
conditionoccurs.
●Checkthesensorand
connections.
●Checkthepatientstatus.
●Demomode
●SpO2Demomode
●param.
●al.area
CARESCAPESpO
2
–Masimoisin
demomode.
●ExitDemomode.Seethe
supplementalinformation
provided.
●Expiringcable
●param.
CARESCAPESpO
2
–Masimohas
detectedthattheinterconnect
cableisnearitsexpirationdate.
●Replacetheinterconnect
cable.
●Expiringprobe
●param.
CARESCAPESpO
2
–Masimo
hasdetectedthatthesensor
oradhesivesensorisnearits
expirationdate.
●Replacethesensoror
adhesivesensor.
●Faultycable
●SpO2faultycable
●param.
●al.area
CARESCAPESpO
2
–Masimo
hasdetectedadefective
interconnectcable.
●Replacetheinterconnect
cable.
●Faultyprobe
●SpO2faultyprobe
●FaultyadhesiveSpO2
probe
●param.
●al.area
Thesensorhasfailed.
Theadhesivesensorhasfailed.
●Replacethesensor.
●Replacetheadhesivesensor.
●IdenticalSpO2device●al.areaYoucanonlyhaveone
CARESCAPESpO
2
,CARESCAPE
SpO
2
–Masimo,orCARESCAPE
SpO
2
–Nellcorinthesystem.
●Removeallbutone
CARESCAPESpO
2
device.
●Incompatiblecable
●IncompatibleSpO2
cable
●param.
●al.area
CARESCAPESpO
2
–Masimohas
detectedthatanincompatible
orunrecognizedinterconnect
cablehasbeenconnected.
●ConnectaMasimo
compatibleinterconnect
cable.