Messages
Message
Location
PossiblecausesSuggestedactions
●RVP1meanhigh/
RVP1meanlowto
RVP2meanhigh/
RVP2meanlow
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●UAC1disconnectto
UAC2disconnect
●Disconnected
●al.area
●param.
Noarterialinvasivepressureis
detected.
●Checkthepatientstatus.
●Checkconnections.
●Ifpressuredropsbecauseof
zeroing,performthezeroing
process.
●UAC1syshigh/UAC
1syslowtoUAC2sys
high/UAC2syslow
●UAC1meanhigh/
UAC1meanlowto
UAC2meanhigh/
UAC2meanlow
●UAC1diahigh/UAC
1dialowtoUAC2dia
high/UAC2dialow
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●UVC1meanhighto
UVC2meanhigh/
UVC1meanlowto
UVC2meanlow
●al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
●Checkthepatientstatus.
●Adjustalarmlimitsif
necessary.
●Zeroed
●param.
Zeroingwassuccessful.●Noactionrequired.
Messageisautomatically
removedafter10seconds.
●Zeroing
●param.
IPchanneliscurrentlybeing
zeroed.
●Noactionrequired.
Messageisautomatically
removedandreplacedwiththe
zeroingresultsaftercompletion.
●Zeroingfailed●al.areaPulsatingwaveformdetected.
Defectivetransducer
Offsetis>150mmHg.
●Openthetransducertoroom
airandzerothechannel.
●Replacethetransducer,open
ittoroomair,andzerothe
channel.
●ZeroICPseparately●al.areaTheICPchannelmustbezeroed
separatelyfromallotherinvasive
pressures.
●Zerothechannelusingthe
Zerooptionfoundunderthe
ICPchannelsetupmenu.
224CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017