Location
PossiblecausesSuggestedactions
RVP1meanhigh/
RVP1meanlowto
RVP2meanhigh/
RVP2meanlow
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
UAC1disconnectto
UAC2disconnect
Disconnected
al.area
param.
Noarterialinvasivepressureis
detected.
Checkthepatientstatus.
Checkconnections.
Ifpressuredropsbecauseof
zeroing,performthezeroing
process.
UAC1syshigh/UAC
1syslowtoUAC2sys
high/UAC2syslow
UAC1meanhigh/
UAC1meanlowto
UAC2meanhigh/
UAC2meanlow
UAC1diahigh/UAC
1dialowtoUAC2dia
high/UAC2dialow
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
UVC1meanhighto
UVC2meanhigh/
UVC1meanlowto
UVC2meanlow
al.areaMeasurementvaluesareequal
tooroutsidethealarmlimits.
Checkthepatientstatus.
Adjustalarmlimitsif
necessary.
Zeroed
param.
Zeroingwassuccessful.Noactionrequired.
Messageisautomatically
removedafter10seconds.
Zeroing
param.
IPchanneliscurrentlybeing
zeroed.
Noactionrequired.
Messageisautomatically
removedandreplacedwiththe
zeroingresultsaftercompletion.
Zeroingfailedal.areaPulsatingwaveformdetected.
Defectivetransducer
Offsetis>150mmHg.
Openthetransducertoroom
airandzerothechannel.
Replacethetransducer,open
ittoroomair,andzerothe
channel.
ZeroICPseparatelyal.areaTheICPchannelmustbezeroed
separatelyfromallotherinvasive
pressures.
Zerothechannelusingthe
Zerooptionfoundunderthe
ICPchannelsetupmenu.