Alarmspecications
AlarmLimitrangeLimitincrementAlarmdelay
●FemVXmeanlow
●ICPXmeanlow
●LAPXmeanlow
●P1meanlowto
P2meanlow
●RAPXmeanlow
●RVPXmeanlow
●UACXsyslow/
UACXdialow/
UACXmeanlow
●UVCXmeanlow
CPPXhigh
-25to320mmHg
(-3.3to42.7kPa)
1mmHg(0.1kPa)
16s(average)(14to17srange)
CPPXlow
-25to320mmHg
(-3.3to42.7kPa)
1mmHg(0.1kPa)
18s(average)(16to20srange)
●PR(Art)Xhigh
●PR(Fem)Xhigh
●PR(UAC)Xhigh
20to300bpm1bpm
7s
●PR(Art)Xlow
●PR(Fem)Xlow
●PR(UAC)Xlow
20to300bpm1bpm
7s
Alarmdelayspecicationsfortemperaturealarms
Thefollowingtableliststhealarmdelaysforphysiologicalalarmsrelatedtothetemperature
measurement.Anyalarmdelaysthatarelessthan5saregivenas<5s.
AlarmLimitrangeLimitincrementAlarmdelay
T1high
T2high
T1low
T2low
10to45°C(50to
113°F)
0.1°C(0.18°F)61s(average)(60to61srange)
AlarmdelayspecicationsforCO
2
alarms
ThefollowingtableliststhealarmdelaysforphysiologicalalarmsrelatedtotheCO
2
measurement.Anyalarmdelaysthatarelessthan5saregivenas<5s.
AlarmLimitrangeLimitincrementAlarmdelay
Apnea(CO2)20s(average)(20srange)
EtCO2high/EtCO2
low
0to13%0.1%
21s(average)(20to23srange)
FiCO2high/FiCO2
low
0.0%to3%0.1%
21s(average)(20to23srange)
268CARESCAPEONE2094480-001