Alarmspecications
AlarmParameterAlarmdelay
Checkdevice
ArtXpressuremalfunction
CPPXpressuremalfunction
CVPXpressuremalfunction
FemXpressuremalfunction
FemVXpressuremalfunction
ICPXpressuremalfunction
LAPXpressuremalfunction
PAXpressuremalfunction
RAPXpressuremalfunction
RVPXpressuremalfunction
UACXpressuremalfunction
UVCXpressuremalfunction
whereX=channelnumber(1or
2)
Invasivepressures
<5s
Checkdevice
SpO2malfunction
SpO
2
<5s
Checkdevice
Temperaturemalfunction
Temperature
8s
CheckNIBP
NIBP<5s
Checksampleline
CO
2
40s(average)(39to40srange)
Checksensor
ServiceCO2sensor
CO
2
<5s
CheckSpO2probe
Checkprobe
SpO
2
<5s
CO2devicefailure:Callservice
CO
2
7s
CO2deviceoverheated
Shuttingdown
CO
2
10s
CO2sensorremoved
Sensorremoved
CO
2
<5s
CO2zerorequired
Zerorequired
CO
2
42s(average)(41to42srange)
Deviceoverheating
MoveCO2devicetoacooler
location
CO
2
6s
Deviceoverheating
MoveECGdevicetoacooler
location
ECG6s
Deviceoverheating
Movepressuredevicetoa
coolerlocation
Invasivepressures
7s
270CARESCAPEONE2094480-001