Alarmspecications
PRIORITIESANDESCALATIONTIMESMESSAGENOTESLOCATION
infolowmediumhigh
●FemXsys
high/FemX
syslow
●FemXmean
high/FemX
meanlow
●FemXdiahigh
/FemXdia
low
Alarmpriorityset
tolow,medium,
orhigh
alarm10s,accordingtoprioritysetting:high,
medium,low
FemVXmean
high/FemVX
meanlow
Alarmpriorityset
toEscalating
alarm
10s30s
10sif
Brady/
TachyorHR
high/low
alarmalso
present
FemVXmean
high/FemVX
meanlow
Alarmpriorityset
tolow,medium,
orhigh
alarm10s,accordingtoprioritysetting:high,
medium,low
ICPXmeanhigh
/ICPXmeanlow
Alarmpriorityset
toEscalating
alarm
10s30s
10sif
Brady/
TachyorHR
high/low
alarmalso
present
ICPXmeanhigh
/ICPXmeanlow
Alarmpriorityset
tolow,medium,
orhigh
alarm10s,accordingtoprioritysetting:high,
medium,low
Identical
pressuredevices
alarm
0s
IP'snotzeroedalarm
0s300s
LAPXmeanhigh
/LAPXmeanlow
Alarmpriorityset
toEscalating
alarm
10s30s
10sif
Brady/
TachyorHR
high/low
alarmalso
present
LAPXmeanhigh
/LAPXmeanlow
Alarmpriorityset
tolow,medium,
orhigh
alarm10s,accordingtoprioritysetting:high,
medium,low
Movepressure
devicetoacooler
location
alarm
0s
P1notzeroedto
P2notzeroed
param.
0s