Defaultsettings
AllowedPrioritiesOtherParametersdefaultsettings
SettingDescription
ED
ICUNICUORPACU
CareUnitSettings>ParameterAlarms>AllowedPrioritiesOtherParameters>Page1orPage2
Apnea(Imped.)Selecttheallowed
alarmprioritiesfor
impedanceapnea
alarms.
Escalating,High
CO2High/Low
Escalating
Allowed
Selectifthe
escalatingalarm
priorityisallowed.
Disabled
RR(Imped.)
High/Low
Selecttheallowed
alarmpriorities
forimpedance
respirationrate
high/lowalarms.
Escalating,High
Temperature
Alarms
Informational
Allowed
Selectifthe
informationallevel
alarmsettingis
allowed.
Disabled
Resp(Imped.)
Meas.Paused
Selecttheallowed
alarmpriorities
forimpedance
respiration
measurement
pausedalarms.
Escalating,High
AlarmDeactivationdefaultsettings
SettingDescription
ED
ICUNICUORPACU
CareUnitSettings>ParameterAlarms>AlarmDeactivation>Page1orPage2
Allowalarm
deactivationfrom
theparameter
menufor:SpO2
ProbeOff
Selectifthe
DeactivateSpO2
ProbeOffoption
isenabledinthe
SpO2menu.
Disabled
Allowalarm
deactivationfrom
theparameter
menufor:Apnea
(CO2)
Selectifthe
DeactivateApnea
Alarmoptionis
enabledintheCO
2
menu.
Disabled
Allowalarm
deactivationwith
theAudioPause
keyfor:ECGLeads
Off
SelectiftheLeads
offalarmcanbe
deactivatedwith
theAudioPause
button.
Disabled
Allowalarm
deactivationwith
theAudioPause
keyfor:Arterial
Disconnect
SelectifArt,
Fem,andUAC
disconnectalarms
canbedeactivated
withtheAudio
Pausebutton.
Disabled
2094480-001CARESCAPEONE329