Defaultsettings
Setting
ED
ICUNICUORPACU
FemSysHigh
Limit/LowLimit
180/80
(Neuroprole:160/90;
Adolescentprole:150/80;
Pediatricprole:130/70;
Infantprole:120/50)
(USonly:Infantprole:100/50)
90/40
(Cardiac,
Full-term,and
Infantproles:
100/40)
(USonly:
Cardiacand
Full-term
proles:
100/40;
Infantprole:
100/50)
180/80
(Neuroprole:160/90;
Adolescentprole:150/80;
Pediatricprole:130/70;Infant
prole:120/50)
(USonly:Infantprole:100/50)
FemSysAlarmOnEnabled
FemMeanHigh
Limit/LowLimit
140/60
(Adolescentprole:110/50;
Pediatricprole:100/50;
Infantprole:80/40)
60/30
(Cardiac,
Full-term,and
Infantproles:
70/40)
(USonly:
Infantprole:
80/40)
140/60
(Adolescentprole:110/50;
Pediatricprole:100/50;
Infantprole:80/40)
FemMeanAlarmOnDisabled
(Adolescent,Pediatric,and
Infantproles:Enabled)
(USonly:Enabled)
EnabledDisabled
(Adolescent,Pediatric,and
Infantproles:Enabled)
(USonly:Enabled)
FemDiaHigh
Limit/LowLimit
100/40
(AdolescentandPediatric
proles:80/40;Infant
prole:70/30)
50/20
(Cardiac,
Full-term,and
Infantproles:
60/20)
(USonly:
Infantprole:
70/20)
100/40
(AdolescentandPediatric
proles:80/40;Infant
prole:70/30)
FemDiaAlarmOnDisabled
(USonly:Enabled)
PR(Fem)High
Limit/LowLimit
150/50
(Neuroprole:120/50;
Pediatricprole:170/70;
Infantprole:180/80)
200/90
(Infantprole:
180/80)
160/40
(Neuroprole:
120/50;
Pediatric
prole:
170/70;
Adolescent
prole:
150/50;
Infantprole:
180/80)
150/50
(Neuroprole:
120/50;
Pediatric
prole:
170/70;
Infantprole:
180/80)
334
CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017