Defaultsettings
SettingDescription
ED
ICUNICUORPACU
LeadAnalysisSelectthe
lead(s)for
ECGanalysis
whena5-,6-,
10-leadcableis
connected.
Multilead
(3-leadcableconnected:Singlelead)
ShowwithHRSelectthe
highest
deviationST
value,PVC
value,asecond
HR,ornothing
todisplayinthe
ECGparameter
window.
Off
ECGGridDisplayagrid
onwaveforms.
Disabled
ECGHRalarmsdefaultsettings
SelectMonitorSetup>MainSetup>ParameterSetup>Page1>ECG>HRAlarms
Arrhythmia>HR/PRtoadjustthesesettingsforthecurrentpatientwhentheHRAlarms
careunitsettingisMultiple.Thefollowingtableliststhefactorydefaultsettings.
SettingDescription
ED
ICUNICUORPACU
HRalarmlimitsSelectthe
high/lowalarm
limits.
150/50
(Neuroprole:120/50;
Pediatricprole:170/70;
Infantprole:180/80)
200/90
(Infant
prole:
180/80)
160/40
(Neuro
prole:
120/50;
Adolescent
prole:
150/50;
Pediatric
prole:
170/70;
Infant
prole:
180/80)
150/50
(Neuro
prole:
120/50;
Pediatric
prole:
170/70;
Infant
prole:
180/80)
HRalarmonEnablealarm
whenlimitsare
violated.
Enabled
(ORsoftwarepackagePerfusionprole:Disabled)
352CARESCAPEONE2094480-001
CARESCAPE ONE User Manual DRAFT 26 October 2017